A20-UF Ultrafine CPC
cpc

A20-UF Ultrafine CPC

Ultrafine particle detection

The Airmodus A20-UF is an ultrafine butanol-based Condensation Particle Counter designed for detecting aerosol particles down to 3 nm. It is an exclusive variant of the A20 CPC platform, calibrated for maximum sensitivity at the ultrafine particle range. Listed on the ACTRIS UF whitelist as a "golden standard reference CPC", it serves both as a standalone total particle counter and as a detector within broader aerosol measurement systems such as dualCPC configurations. The A20-UF features a touchscreen interface, advanced signal diagnostics, and the BlockSafe inlet system in an 8.5 kg package consuming only 100 W steady state.

Request a Quote

Specifications

Rohs YES
Weight 8.5 kg (instrument only); 9.35 kg with bottle kit + PSU
En 16976 YES
Fittings 6 mm push-fit inlet and vacuum
Power Max 160 W
Ce Marking YES
Dimensions 260 x 230 x 400 mm (H x W x D)
Cutoff Dp50 2.8-3.8 nm (silver particles, see calibration cert) / ~4 nm PG
Cutoff Dp90 < 10 nm
Connectivity USB type B, Ethernet RJ45, RS-232, Analog BNC 0-10V, Pulse BNC
Data Logging 1 Hz (default), 10 Hz, raw pulse
False Counts < 0.001 #/cm3
Operating Rh 0-95% non-condensing
Actris Status UF Whitelist
Working Fluid n-Butanol (>99.5%) standard; Propylene Glycol available
Operating Temp 15-35 C
Dew Point Limit < 10 C
External Vacuum 100-400 mbar at NTP
Dualcpc Compliant YES
Response Time T95 < 1.15 s
Aerosol Inlet Flow 1.05 lpm +/- 5%
Operating Pressure 75-105 kPa
Power Steady State 100 W
Aerosol Sample Flow 1.05 lpm
Concentration Range 0 - 150,000 #/cm3
Particle Size Range 3 nm - 2.5 um (butanol) / 4 nm - 2.5 um (PG)
Measurement Fields [{'kind': 'numeric', 'name': 'concentration', 'unit': 'p/cm3', 'label': 'Particle concentration'}, {'kind': 'numeric', 'name': 'averaging_time', 'unit': 's', 'label': 'Averaging time'}, {'kind': 'numeric', 'name': 'pulse_duration', 'unit': 'us', 'label': 'Pulse duration'}, {'kind': 'numeric', 'name': 'dead_time', 'unit': '', 'label': 'Dead time (multilogger only)'}, {'kind': 'numeric', 'name': 'liquid_level', 'unit': '%', 'label': 'Liquid level'}, {'kind': 'numeric', 'name': 'number_of_pulses', 'unit': '', 'label': 'Pulse count'}, {'kind': 'numeric', 'name': 'pulses', 'unit': '', 'label': 'Pulses (THRS1)'}, {'kind': 'numeric', 'name': 'pulses_thrs2', 'unit': '', 'label': 'Pulses above threshold 2'}, {'kind': 'numeric', 'name': 'dc_compensation_on', 'unit': '', 'label': 'DC compensation enabled'}, {'kind': 'numeric', 'name': 'total_errors', 'unit': '', 'label': 'Total errors (multilogger only)'}, {'kind': 'status', 'name': 'status_hex', 'unit': '', 'label': 'Status word'}, {'kind': 'numeric', 'name': 'temp_cabin', 'unit': 'C', 'label': 'Cabin temperature'}, {'kind': 'numeric', 'name': 'temp_condenser', 'unit': 'C', 'label': 'Condenser temperature'}, {'kind': 'numeric', 'name': 'temp_optics', 'unit': 'C', 'label': 'Optics temperature'}, {'kind': 'numeric', 'name': 'temp_saturator', 'unit': 'C', 'label': 'Saturator temperature'}, {'kind': 'numeric', 'name': 'pres_cabin', 'unit': 'hPa', 'label': 'Cabin pressure'}, {'kind': 'numeric', 'name': 'pres_critical_orifice', 'unit': 'hPa', 'label': 'Critical orifice pressure'}, {'kind': 'numeric', 'name': 'pres_inlet', 'unit': 'hPa', 'label': 'Inlet pressure'}, {'kind': 'numeric', 'name': 'pres_nozzle', 'unit': 'hPa', 'label': 'Nozzle pressure'}, {'kind': 'numeric', 'name': 'pres_wrem', 'unit': 'hPa', 'label': 'Water removal pressure'}, {'kind': 'numeric', 'name': 'pulse_ratio', 'unit': '', 'label': 'Pulse quality ratio (THRS2/THRS1)'}]