PSM 2.0 Retrofit Kit
psm

PSM 2.0 Retrofit Kit

Upgrade existing PSM to 2.0 specifications

Retrofit upgrade kit that brings older Airmodus PSM hardware up to PSM 2.0 firmware and electronics specifications, including the full 1-12 nm scanning range. Shares the PSM 2.0 22-field measurement format with one hardware-level difference: vacuum flow is controlled by a passive critical orifice (no vacuum MFC), so the vacuum_flow field is always reported as 0.0.

Request a Quote

Specifications

Sample Rh 0-95% non-condensing (above 40% RH dilute with dry air via AFC; Nafion dryer not recommended due to diffusional losses sub-12 nm)
Connectivity RS-232, USB type B, Analog out BNC 0-10V (for DMA/ion filter control)
Operating Rh 0-95% non-condensing
Flow Rate Lpm Aerosol sample 2.5 lpm; sample to detector CPC 1 lpm; saturator 0.1-1.3 lpm (scanning)
Working Fluid Diethylene Glycol (DEG, >99%)
Operating Temp 15-30 C
External Vacuum 100-350 mbar at NTP
Sample Pressure 90-105 kPa
Response Time T95 < 2 s
Operating Pressure 90-105 kPa
Vacuum Flow Control Critical orifice (passive, not measured) - differs from PSM 2.0 which uses a vacuum MFC
Particle Size Range Nm 1-12 nm (Dp50 selectable 1.19-12 nm, NiCr-equivalent activation diameter)
External Compressed Air 1.5-2.5 bar at NTP; particle/oil/water free; dew point < 0 C; max 3.0 bar
Measurement Fields [{'kind': 'numeric', 'name': 'excess_flow', 'unit': 'L/min', 'label': 'Excess flow'}, {'kind': 'numeric', 'name': 'liquid_errors', 'unit': '', 'label': 'Liquid errors'}, {'kind': 'status', 'name': 'note_hex', 'unit': '', 'label': 'Note word'}, {'kind': 'numeric', 'name': 'poly_correction', 'unit': '', 'label': 'Polynomial correction'}, {'kind': 'numeric', 'name': 'pres_critical_orifice', 'unit': 'hPa', 'label': 'Critical orifice pressure'}, {'kind': 'numeric', 'name': 'pres_inlet', 'unit': 'hPa', 'label': 'Inlet pressure'}, {'kind': 'numeric', 'name': 'pres_inlet_saturator', 'unit': 'hPa', 'label': 'Inlet-Saturator pressure'}, {'kind': 'numeric', 'name': 'pres_saturator_excess', 'unit': 'hPa', 'label': 'Saturator-Excess pressure'}, {'kind': 'numeric', 'name': 'sat_flow_setpoint', 'unit': 'L/min', 'label': 'Saturator flow setpoint'}, {'kind': 'numeric', 'name': 'saturator_flow', 'unit': 'L/min', 'label': 'Saturator flow'}, {'kind': 'status', 'name': 'scan_status', 'unit': '', 'label': 'Scan status'}, {'kind': 'status', 'name': 'status_hex', 'unit': '', 'label': 'Status word'}, {'kind': 'numeric', 'name': 'temp_cabin', 'unit': 'C', 'label': 'Cabin temperature'}, {'kind': 'numeric', 'name': 'temp_drainage', 'unit': 'C', 'label': 'Drainage temperature'}, {'kind': 'numeric', 'name': 'temp_growth_tube', 'unit': 'C', 'label': 'Growth tube temperature'}, {'kind': 'numeric', 'name': 'temp_heater', 'unit': 'C', 'label': 'Heater temperature'}, {'kind': 'numeric', 'name': 'temp_inlet', 'unit': 'C', 'label': 'Inlet temperature'}, {'kind': 'numeric', 'name': 'temp_saturator', 'unit': 'C', 'label': 'Saturator temperature'}, {'kind': 'numeric', 'name': 'total_errors', 'unit': '', 'label': 'Total errors'}, {'kind': 'numeric', 'name': 'vacuum_flow', 'unit': 'L/min', 'label': 'Vacuum flow'}, {'kind': 'numeric', 'name': 'temp_mfc', 'unit': 'C', 'label': 'MFC temperature'}]